Operation 4.5.2 hold/transfer outputs, 3 release outputs – Hach-Lange SC 60 User Manual
Page 37
Advertising
33
Operation
4.5.2 Hold/Transfer Outputs
During normal measurement operation, the analog outputs can be
held at the last measured value or transferred to a preset value.
During calibration, analog outputs can remain active, held, or
transferred to a preset mA value.
4.5.2.1 Hold/Transfer Outputs until Release
Step
Select
Menu Level/Instructions
Confirm
1
MAIN MENU
—
2
TEST/MAINT
3
HOLD OUTPUTS
4
SET OUTMODE
5
Select HOLD OUTPUTS or XFER OUTPUTS
6
SET CHANNELS
7
Select the connected sensor
—
8
ACTIVATION
9
Select LAUNCH
10
Main Menu or Main Measurement Screen
reading
will flash
4.5.3 Release Outputs
Step
Select
Menu Level
Confirm
1
MAIN MENU
—
2
TEST/MAINT
3
HOLD OUTPUTS
4
ACTIVATION
5
RELEASE
6
Main Menu or Main Measurement Screen
—