Operation 4.5.2 hold/transfer outputs, 3 release outputs – Hach-Lange SC 60 User Manual

Page 37

Advertising
background image

33

Operation

4.5.2 Hold/Transfer Outputs

During normal measurement operation, the analog outputs can be
held at the last measured value or transferred to a preset value.

During calibration, analog outputs can remain active, held, or
transferred to a preset mA value.

4.5.2.1 Hold/Transfer Outputs until Release

Step

Select

Menu Level/Instructions

Confirm

1

MAIN MENU

2

TEST/MAINT

3

HOLD OUTPUTS

4

SET OUTMODE

5

Select HOLD OUTPUTS or XFER OUTPUTS

6

SET CHANNELS

7

Select the connected sensor

8

ACTIVATION

9

Select LAUNCH

10

Main Menu or Main Measurement Screen

reading

will flash

4.5.3 Release Outputs

Step

Select

Menu Level

Confirm

1

MAIN MENU

2

TEST/MAINT

3

HOLD OUTPUTS

4

ACTIVATION

5

RELEASE

6

Main Menu or Main Measurement Screen