A.2 sc60 modbus registers (continued) – Hach-Lange SC 60 User Manual

Page 55

Advertising
background image

51

Setup/Relay 1

Measurement
Select

40052

Unsigned
Integer

1

R/W

Selects measurement on the Sensor
(0 = Meas1; 3 = Meas4)

Setup/Relay 1

Type

40053

Unsigned
Integer

1

R/W

Selects the relay type
(0 = Alarm; 1 = Control;
2 = Status; 3 = Timer; 4 = Event)

Setup/Relay 1

Transfer Setting 40054

Unsigned
Integer

1

R/W

Selects the transfer value for the relays
(0 = De-energized; 1 = Energized)

Setup/Relay
1/Alarm

High Alarm

40055

Float

2

R/W

Sets the high alarm setpoint

Setup/Relay
1/Alarm

Low Alarm

40057

Float

2

R/W

Sets the low alarm setpoint

Setup/Relay
1/Alarm

High Deadband

40059

Float

2

R/W

Sets the high alarm deadband

Setup/Relay
1/Alarm

Low Deadband

40061

Float

2

R/W

Sets the low alarm deadband

Setup/Relay
1/Alarm

On Delay

40063

Unsigned
Integer

1

R/W

Sets the on delay time

Setup/Relay
1/Alarm

Off Delay

40064

Unsigned
Integer

1

R/W

Sets the off delay time

Setup/Relay
1/Control

Setpoint

40065

Float

2

R/W

Sets the controller setpoint

Setup/Relay
1/Control

Phase

40067

Unsigned
Integer

1

R/W

Sets the controller phase
(0 = Low; 1 = High)

Setup/Relay
1/Control

Deadband

40068

Float

2

R/W

Sets the controller deadband

Setup/Relay
1/Control

Overfeed Timer

40070

Unsigned
Integer

1

R/W

Sets the overfeed timer value (mins)

Setup/Relay
1/Control

On Delay

40071

Unsigned
Integer

1

R/W

Sets the on delay time (sec)

Setup/Relay
1/Control

Off Delay

40072

Unsigned
Integer

1

R/W

Sets the off delay time (sec)

Setup/Relay
1/Control

Reset Overfeed
Timer

40073

Unsigned
Integer

1

R/W

Resets the overfeed timer

Setup/Relay
1/Event

Setpoint

40074

Float

2

R/W

Sets the event setpoint

Setup/Relay
1/Event

Phase

40076

Unsigned
Integer

1

R/W

Sets the event phase (0 = Low; 1 = High)

Setup/Relay
1/Event

Deadband

40077

Float

2

R/W

Sets the event deadband

Setup/Relay
1/Event

On Max Time

40079

Unsigned
Integer

1

R/W

Sets the max on time (mins)

Setup/Relay
1/Event

On Min Time

40080

Unsigned
Integer

1

R/W

Sets the min on time (mins)

Setup/Relay
1/Event

Off Max Time

40081

Unsigned
Integer

1

R/W

Sets the max off time (mins)

Setup/Relay
1/Event

Off Min Time

40082

Unsigned
Integer

1

R/W

Sets the min off time (mins)

A.2 sc60 Modbus Registers (continued)