A.2 sc60 modbus registers (continued) – Hach-Lange SC 60 User Manual

Page 59

Advertising
background image

55

Setup/Relay
3/Timer

Duration Time

40166

Unsigned
Integer

1

R/W

Sets the timer on duration time (sec)

Setup/Relay
3/Timer

Period Time

40167

Unsigned
Integer

1

R/W

Sets the period between timer on events
(mins)

Setup/Relay
3/Timer

Off Delay

40168

Unsigned
Integer

1

R/W

Sets the time the affected Sensor outputs
are held/transferred after the timer turns
off (sec)

Setup/Relay
3/Status

Level

40169

Unsigned
Integer

1

R/W

Sets the status level which will trigger
the relay

Comm/Net Card

Mode

40170

Unsigned
Integer

1

R/W

Sets the Modbus mode
(0 = RTU; 1 = ASCII)

Comm/Net Card

Baud

40171

Unsigned
Integer

1

R/W

Sets the Modbus baud rate (0 = 9600; 1 =
19200; 2 = 38400; 3 = 57600;
4 = 115200)

Comm/Net Card

Stop Bits

40172

Unsigned
Integer

1

R/W

Sets the number of stop bits (1.2)

Comm/Net Card

Data Order

40173

Unsigned
Integer

1

R/W

Sets the register data order for floats
(0 = Normal; 1 = Reversed)

Comm/Net Card

Min Response
Time

40174

Unsigned
Integer

1

R/W

Sets the minimum response time
(0 to 30 sec)

Comm/Net Card

Max Response
Time

40175

Unsigned
Integer

1

R/W

Sets the maximum response time
(100 to 1000 sec)

Comm/Net
Card/Addresses

sc60

40176

Unsigned
Integer

1

R/W

Sets the sc60 Modbus Address

Comm/Net
Card/Addresses

Sensor 1

40177

Unsigned
Integer

1

R/W

Sets the Sensor 1 Modbus Address

Comm/Net
Card/Addresses

Sensor 2

40178

Unsigned
Integer

1

R/W

Sets the Sensor 2 Modbus Address

Comm/Net
Card/Stats

Good
Messages

40179

Unsigned
Integer

2

R/W

Number of good messages

Comm/Net
Card/Stats

Bad Messages

40181

Unsigned
Integer

2

R/W

Number of failed messages

Comm/Net
Card/Stats

% Good Mesg

40183

Float

2

R/W

% of good messages

Comm/Service
Port

Mode

40185

Unsigned
Integer

1

R/W

Sets the Modbus mode (0 = RTU;
1 = ASCII)

Comm/Service
Port

Baud

40186

Unsigned
Integer

1

R/W

Sets the Modbus baud rate (0 = 9600;
1 = 19200; 2 = 38400; 3 = 57600;
4 = 115200)

Comm/Service
Port

Stop Bits

40187

Unsigned
Integer

1

R/W

Sets the number of stop bits (1.2)

Comm/Service
Port

Data Order

40188

Unsigned
Integer

1

R/W

Sets the register data order for floats
(0 = Normal; 1 = Reversed)

Comm/Service
Port

Min Response
Time

40189

Unsigned
Integer

1

R/W

Sets the minimum response time
(0 to 30 sec)

Comm/Service
Port

Max Response
Time

40190

Unsigned
Integer

1

R/W

Sets the maximum response time
(100 to 1000 sec)

Comm/Service
Port/Addresses

sc60

40191

Unsigned
Integer

1

R/W

Sets the sc60 Modbus Address

A.2 sc60 Modbus Registers (continued)

Group Name

Tag Name

Register #

Data

Type

Length

R/W

Description