Installation – Hach-Lange SC 60 User Manual

Page 19

Advertising
background image

15

Installation

The DC power source that supplies power to the 24 VDC powered
sc60 must maintain voltage regulation within the specified 24 VDC
–15 % +20 % voltage limits. The DC power source must also
provide adequate protection against surges and line transients.

Figure 11 Proper Wire Preparation and Insertion

1

Strip ¼-inch of insulation.

2

Seat insulation against connector with no bare wire exposed.

Table 2: AC Power Wiring Information (sc60 AC powered model only)

Terminal Number

Terminal Description

Wire Color Code for North America

Wire Color Code for Europe

1

Hot (L1)

Black

Brown

2

Neutral (N)

White

Blue

3

Protective Earth (PE)

Green

Green w/yellow tracer

Table 3: DC Power Wiring Information (sc60 24 VDC powered model only)

Terminal Number

Terminal Description

Wire Color Code for North America

Wire Color Code for Europe

1

+24 V dc

Red

Red

2

24 V dc return

Black

Black

3

Protective Earth (PE)

Green

Green w/yellow tracer