Installation – Hach-Lange SC 60 User Manual
Page 19
15
Installation
The DC power source that supplies power to the 24 VDC powered
sc60 must maintain voltage regulation within the specified 24 VDC
–15 % +20 % voltage limits. The DC power source must also
provide adequate protection against surges and line transients.
Figure 11 Proper Wire Preparation and Insertion
1
Strip ¼-inch of insulation.
2
Seat insulation against connector with no bare wire exposed.
Table 2: AC Power Wiring Information (sc60 AC powered model only)
Terminal Number
Terminal Description
Wire Color Code for North America
Wire Color Code for Europe
1
Hot (L1)
Black
Brown
2
Neutral (N)
White
Blue
3
Protective Earth (PE)
Green
Green w/yellow tracer
Table 3: DC Power Wiring Information (sc60 24 VDC powered model only)
Terminal Number
Terminal Description
Wire Color Code for North America
Wire Color Code for Europe
1
+24 V dc
Red
Red
2
24 V dc return
Black
Black
3
Protective Earth (PE)
Green
Green w/yellow tracer
1