Specifications – SKC Limited AirChek 3000 Deluxe Pump Operating Instructions User Manual

Page 11

Advertising
background image

www.skcltd.com

210-3311M Issue E

Page 9

Specifications

Pump Features

Flow fault function

Indicates flow fault due to obstructed tubing or excessive filter loading. Shuts
the pump down and enters ‘HOLD’ mode if the condition persists for longer
than 15 seconds. Automatically attempts to restart every five minutes (up to
a maximum of 10 restarts), until the flow fault condition clears in which case
normal running will resume.

Programmable run time

Sample run time programmable in minutes, via keypad and LCD screen.
Pump automatically shuts down at end of sample and retains run time data.

Programmable delayed
start time

Sample start delay time programmable in hours and minutes, via keypad and
LCD screen.

Particulate trap

Built in replaceable filter to trap particles that would otherwise contaminate
the pump mechanism.

Automatic air flow rate
calibration

With optional Defender primary calibrator and CalChek Communicator cable.
Enables automatic calibration of the sample air flow rate to the required level.

PC Connectivity

With optional DataTrac USB adapter cable and PC software. Enables
programming of timed sample runs, delayed starts and intermittent sampling.
Enables retrieval of pump run-time data and history to the PC.