User Interface Level One

10. At the end of a successful full calibration the pump will automatically switch to

HOLD mode. The full calibration data is written into the pump memory when the
pump switches to SLEEP mode, therefore leave the pump in HOLD mode for five
minutes after which it will switch to SLEEP mode.

11. If a problem was encountered during the full calibration, an error code will be

displayed. Refer to the CalChek Error Code Table on pages 42 and 43. Press the

button to clear the error code screen and revert to HOLD mode.

The full calibration process can be aborted at any stage by pressing [

]. The pump will automatically switch

to HOLD mode.

The full calibration data can be viewed and printed using the optional DataTrac Interface and Software and a
PC. Access the DataTrac Software Pump Manager window, click on the View menu and select the Calibration
Info option. A new window will display the calibration results, pump serial number and date of the last full
calibration. A button on this window allows the data to be printed. The printed report also includes the pump
internal software version, the date printed and a validation code.

To ensure that the printed calibration data has not been tampered with, in the DataTrac Software click on the
Tools menu and select Confirm Validation Code. Enter the data from the printed report, including the validation
code. The date is entered in the format mmm dd, yyyy (e.g. Mar 12, 2011). The DataTrac Software will indicate
whether the information is completely valid or if a parameter has been changed.

Note: Clearing the pump history from memory will not clear the full calibration data. This data can only be
cleared by performing another full calibration or by obtaining more than 36 pump history records.