User interface level one – SKC Limited AirChek 3000 Deluxe Pump Operating Instructions User Manual

Page 24

Advertising
background image

Page 22

210-3311M Issue E

www.skcltd.com

User Interface Level One

1) Setting and Manual Calibration of the Pump Flow Rate

1. Set up the sample train as specified in the sampling method and connect to the pump air inlet port.

Connect a calibrated flow meter such as a Defender primary flow calibrator to the inlet of the sample
train.

2. With the pump in HOLD mode, press [

] to switch to RUN mode, and allow the pump to run for at

least 5 minutes before commencing setting and calibration of the flow rate.

3. Enter User Interface Level One as detailed in Getting Started 7) on page 19.

4. The first User Interface screen displayed is the flow setting screen, with ‘

SET

flashing.

5. Use the and buttons to increase or decrease the flow setting, in steps of 10

ml/min, to the required flow rate on the LCD screen.

6. When the required flow rate is set press the button to display the flow rate

adjust screen, with ‘

ADJ’ flashing.

7. Set the calibrated flow meter to take continuous readings, and observe the

indicated flow rate. Use the and buttons to adjust the pump flow rate up or
down in steps of 10 ml/min, until the calibrated flow meter indicates the required
flow rate to within ±10 ml/min.