General Troubleshooting

Possible Fault

Corrective Action

Battery pack will not charge

• Check the battery charger by trying it with a different battery pack.

Replace the battery charger if required.

• Check the battery pack by connecting it to a known good battery

charger. If the battery pack still will not charge replace the battery pack.

Pump will not operate

• Check for faulty battery pack by trying a known good battery pack.

Replace the battery pack if required.

Pump flow faults continuously

• Sample media back pressure too high. Try a lower flow rate and/or a

less restrictive sample media if the sampling method being used allows
this.

• Pump particulate filter is blocked (appears black). Replace the

particulate filter.

• Tubing blocked or crimped. Replace tubing.

Pump cannot achieve required
flow rate

• Battery pack voltage low. Fully charge the battery pack.
• Sample media back pressure too high. Try a lower flow rate and/or a

less restrictive sample media if the sampling method being used allows
this.

• Inlet filter cover / hosetail not screwed on tightly enough. Ensure filter

cover is tightly fitted.

• Pump mechanism leaking. Contact SKC Ltd customer services for

assistance.