Page 34

210-3311M Issue E

www.skcltd.com

5) Setting a Delayed Start Time

The delayed start time is the 12 hour clock time at which the pump will automatically switch from HOLD
mode to RUN mode and commence sampling. The delayed start time set has no AM or PM designation, and
the actual start time will be the next occurence of this time in the 12 hour clock, either in the AM or PM. The
delayed start time must be set in conjunction with a sample run time.

1. On the pump enter User Interface Level Two as detailed in Getting Started 7) on page 19.
2. Enter a sample run time as detailed in User Interface Level Two 3) on page 31. This is required to set a

delayed start time.

3. Press the button until the 12/24 hour clock setting screen is displayed.

4. Use the and buttons to select the delayed start option, with ‘dELA’ flashing

on the LCD screen.

5. Press the button once to display the delayed start time screen, with ‘

SET’ and

the current delayed start time hours value flashing.

6. Use the and buttons to enter the required delayed start time hours value.

User Interface Level Two