CalChek Calibration Error Code Troubleshooting

Error Code Problem

Corrective Action

E 46

Full calibration - data analysis error.

Repeat the full calibration.
If this does not resolve the problem contact
SKC Ltd customer services for assistance.

E 47

Full calibration - Battery charge level is too
low. Less than two bars displayed on the
battery status indicator.

Fully charge the battery before repeating the
full calibration.

E 48

Single point calibration - Pump could not
achieve a successful calibration within 5
Defender flow readings.

Repeat the calibration.
If the problem repeats perform a full calibration.