User Interface Level One

2. With the pump in HOLD mode, press [

] to switch to RUN mode, and allow the pump to run for at

least 5 minutes before commencing setting and calibration of the flow rate.

3. On the Defender calibrator, ensure that the ‘Data Port’ parameter on the Setup - Preferences screen is

set to the ‘SKC’ option. Set the Defender to take continuous readings. Refer to the Defender instructions
for further information.

4. On the pump enter User Interface Level One as detailed in Getting Started 7) on page 19.

5. The first User Interface screen displayed is the flow setting screen, with ‘

SET

flashing.

6. Use the and buttons to increase or decrease the flow setting, in steps of 10

ml/min, to the required flow rate on the LCD screen.

7. When the required flow rate is set press the button until the ‘CAL’ screen is

displayed.

8. Press [

] to initiate an automatic single point flow calibration.