5um dust particulates – SKC Limited EPAM-5000 User Manual
Page 29
2.5um Dust Particulates
Note: 2.5-µm size select is the default setting of the EPAM-5000.
Follow the steps in the table below to select PM2.5 dust particulates.
Step Action
1
Select Special Functions from the Main Menu.
2
Select System Options.
3
Select Extended Options.
4
Select Size Select.
5
Select 2.5.
Result: The Main Menu is displayed.
6
Insert the inlet into the sensor head of the EPAM-5000.
Note: If also collecting concurrent 47mm EPA FRM Style filter
samples place a clean gravimetric filter in the filter cassette. The
Flow Rate should be checked each time a new gravimetric filter is
used. See page 5-5 for information on checking the flow rate.
7
Attach the filter cassette holder to the sensor of the EPAM-5000.
8
Turn to page 3-7 and follow the instructions to Manual-Zero the
EPAM-5000.
3 - 4