5um dust particulates – SKC Limited EPAM-5000 User Manual

Page 29

Advertising
background image

2.5um Dust Particulates

Note: 2.5-µm size select is the default setting of the EPAM-5000.


Follow the steps in the table below to select PM2.5 dust particulates.

Step Action

1

Select Special Functions from the Main Menu.

2

Select System Options.

3

Select Extended Options.

4

Select Size Select.

5

Select 2.5.

Result:
The Main Menu is displayed.

6

Insert the inlet into the sensor head of the EPAM-5000.

Note: If also collecting concurrent 47mm EPA FRM Style filter
samples place a clean gravimetric filter in the filter cassette. The
Flow Rate should be checked each time a new gravimetric filter is
used. See page 5-5 for information on checking the flow rate.

7

Attach the filter cassette holder to the sensor of the EPAM-5000.

8

Turn to page 3-7 and follow the instructions to Manual-Zero the
EPAM-5000.










3 - 4

Advertising