Adjusting the Flow Rate

Introduction

The flow rate must be adjusted when it does not equal 4.0 LPM.

Adjusting the
flow rate

Follow the steps in the table below to adjust the flow rate.

1

Locate the adjustment screw on the front of the EPAM-5000.

2

Use the flow adjustment screw to adjust the flow rate.

To...

Turn the adjustment screw...

Decrease the flow rate

Counterclockwise

Increase the flow rate

Clockwise

3

Record the Flow Rate.

4

Detach the airflow calibrator and calibration airflow tubing.

5

Select enter to terminate battery status sensor and return to main
menu.