Battery maintenance – SKC Limited EPAM-5000 User Manual

Page 67

Advertising
background image

5 - 9

Battery Maintenance,

Continued

Recharging the
battery

Follow the steps in the table below to recharge the battery using the supplied
charger.

Note: If the battery is low the sampling process will terminate and the low
battery screen will display.

Step Action

1

Plug the battery charger into an electrical outlet.

2

Plug the battery charger into the battery charge jack on the top
panel of EPAM-5000 Battery.

Results:

• The battery charge begins. LED indicator should be Red.
• Unit must be off or the battery must be removed from

instrument for LED to be Red and charge battery.

• When LED is green battery is fully charged.

Note: Recharging time is approximately 22 hours when the
instrument is not operating, and approximately 24 hours while the
monitor is sampling.

CAUTION:

Do not charge in a hazardous environment. Use only

the EDC approved charger designed for the EPAM-5000.

Continued on next page