Applying a correction factor – SKC Limited EPAM-5000 User Manual

Page 57

Advertising
background image

Applying a Correction Factor

Introduction

A correction factor can be applied to the data collected with the unit to
account for variances in gravimetric readings.

Calculating a
correction
factor

The correction factor is calculated by dividing the Gravimetric reading by the
unit reading.

Applying a
correction
factor

Follow the steps in the table below to apply a correction factor to all data
points in the current project folder.

Step Action

1

Select the 2

nd

Scale= with a box where you can type in your scale

factor.

2

Type in the Scale factor.

3

After the scale factor is entered press enter.

Result: All data points in the project folder have been multiplied
by the correction factor.

Removing the
correction
factor

Follow the steps in the table below to remove the correction factor from the
data points in the project folder.

Step Action

1

Select the 1

st

Scale= under the Dataset Scale Factor.


Result:
Data points should return to original state.











4- 18

Advertising