Manual-zero – SKC Limited EPAM-5000 User Manual

Page 32

Advertising
background image

Manual-Zero

Introduction

Manual-Zero sets the measurement baseline of the EPAM-5000 to zero
mg/m

3

. The Manual-Zero check should take place prior to beginning a new

set of measurements.

Note: If using the Auto-Zero setting (default) the EPAM-5000 automatically
zeroes baseline every 30 minutes.

Manual-Zero

Follow the steps in the table below to Manual-Zero the EPAM-5000.

Note: The battery should be fully charged. Check battery status as described
in Chapter 5 Section 8.

Step Action

1

Be sure the appropriate sampling inlet is attached to the sensor
inlet of the EPAM-5000 using the table below.

If sampling…

Then insert the…

PM1.0 Particulates

1.0 impactor jet

PM2.5 Particulates

2.5 impactor

PM10 Particulates

10 impactor jet

TSP Particulates

Impactor sleeve only

Continued on next page

3 - 7