Manual-zero – SKC Limited EPAM-5000 User Manual

Page 33

Advertising
background image

Manual-Zero,

Continued

3

Select Special Functions from the Main Menu.

4

Select System Options.

5

Select Extended Options.

6

Select Calibration.

Result: Screen appears with manual zero at the top.

7

Select Manual-Zero.
Select Manual-Zero again.

Note: Wait approximately 99 seconds. The unit automatically
executes the steps necessary to reestablish the baseline.

Result: The Main Menu is displayed when the manual-zero
process is complete.

Notes:
The EPAM-5000 has an auto zero purging feature that
automatically purges the sensor and performs an auto zero to
reestablish the baseline approximately every 30 minutes. An X
inside a box appears in right hand corner of the display when the
instrument is auto zeroing.