5 - 2

Checking the Calibration Span

Introduction

The Span Reference Insert Part Number CS-105 is a light scattering device
that provides a constant value (termed a “k” factor).

The Span Reference should be used as a reference to check factory calibration
of the EPAM-5000 Monitor.

When to check
the calibration
span

The calibration span should be checked under the following conditions:
• Once a month with normal usage. (Normal use is twice a week).
• If the EPAM-5000 is dropped or otherwise damaged.
• The first time you use the unit to double check the factory calibration.

Note: The EPAM-5000 can be sent to SKC. Annually recalibration
recommended in accordance with ISO and NIST procedures.

Conditions

The following conditions must be met before checking the calibration span.

Condition...

For further information

see page...

The Sensor Optics must be clean.

5-11

The Environment must be clean.

---

The Battery must be fully charged.

5-7

Figure 5-1. Inserting CS-105 into SKC EPAM-5000 Sensor.

Continued on next page