5.26.2 PCIe Gen3 Auto Calibration

In order for auto calibration at Gen3 to work, lane polarity has to be set manually. In 
general, follow the steps below to find the polarity and set it up properly in the calibration 
dialog.

1. Setup the analyzer to record the training sequence with (TS2, Switch to G3 or 

InitFC1), at Gen1 or Gen2 (use trigger as TS2 or Speed Switch), using the Auto‐
Configure Lane Polarity recording option. This will allow the analyzer to calculate/
determine the polarity and lane reversal settings of the link.

Figure 5.56: Simple Mode Tab

2. Drive the link to Detect or Recovery by powering up or booting the system.

3. After the recording is complete, click the Dashboard View 

 icon to open the 

Dashboard dialog and verify the polarity calculated by the analyzer (see 

Figure 5.57 

on page 106

).