Summit T3‐8 PCI Express Multi‐Lane Protocol Analyzer User Manual

107

Probe Settings

 Teledyne LeCroy

7. Select Recording Options and click the Probe Settings tab. 
8. Click Auto Configure to display the Calibrate SummitT3‐8 Dialog.

Manually set the polarity for each lane in this dialog with the polarity settings in the 
Link Dashboard. Link Width, Speed and Polarity must be set the same as the current 
state of the link. Checking the Quick Calibration option will sweep through the CTLE 
setting only. If unchecked it will sweep through both CTLE and DFE. Using Detect Idle 
Errors will increase the calibration time but it will improve/optimize calibration 
results.

Figure 5.59: Calibrate Summit T3-8 Dialog

9. Click Start Calibration 

On completion of calibration, close the Auto Configure dialog.
It may take 10‐20 minutes for x8 and upwards 45 minutes to calibrate an X16 link for 
both directions with Detect Idle Errors checked and Quick Calibration unchecked.

10. You can now start recording at Gen3 speed using the Recording Option Speed set to 

Gen3 or Auto.