Vestil PM-SCALE User Manual

Page 18

Advertising
background image

10

F1.3 – Calibration

F1.3.1 GEO Adjustment

Selection:

0…31 (Factory default16)


F1.3.2 – Linearity

Selection:

On

- Enabled

OFF - Disabled (Factory default)

F1.3.3 – Calibration
z If F1.3.2 is set to On

[E SCL] Empty scale.
Remove any load from scale platform, then press Enter to confirm. Terminal will
count down from[10 CAL] to [ 0 CAL].

[FULL Ld] Add full load.

Add load on the scale platform, then press Enter.
[000000] Input weight number, then press Enter to confirm. Terminal will count
down from[10 CAL] to [ 0 CAL].

[ donE ] Complete.


z If F1.3.2 is set to OFF

[E SCL] Empty scale.
Remove any load from scale platform, then press Enter to confirm. Terminal will
count down from[10 CAL] to [ 0 CAL].
[Add Ld]Add middle load

Add load on the scale platform, then press Enter.
[000000] Input weight number, then press Enter to confirm. Terminal will count
down from[10 CAL] to [ 0 CAL].

[FULL Ld] Add full load.

Add load on the scale platform, then press Enter.
[000000] Input weight number, then press Enter to confirm. Terminal will count
down from[10 CAL] to [ 0 CAL].

[ donE ] Complete.


F1.4 – Zero Function

F1.4.1 – AZM

Selection:

OFF,0.5d(Factory default),1d,3d


F1.4.2 – Power up Zero

Selection:

OFF,2%,10%(Factory default),20%


F1.4.3 – Pushbutton Zero

Selection:

OFF,2%(Factory default),10%,20%

If F1.1 is set to OIML, Then F1.4.2 will set to 10%, and F1.4.3 will set to 2%.


F1.5 – Tare Function