<5. Display and Operations while Connected to GC1000 Mark II>

5-7

IM 11B06C01-01E

n

Chromatogram Display Section

The chromatogram display section is a part of the Analyzer Operation window that displays the

latest chromatogram for every analysis cycle.
It displays one chromatogram if there is one detector or two chromatograms if there are two

detectors. It also displays the temperature pattern of the programmed temperature oven or

isothermal oven. In addition, this section displays one pressure graph for one EPC or two

pressure graphs for two EPCs.
Because the current elapsed time is indicated by dotted lines, which timing and where the current

location is in the analysis cycle is recognizable at a glance.

SEE ALSO

For the Chromatogram Display window, see Chapter 5.

l

Window elements

F5-2-4.ai

Chromatogram

Temperature

Pressure

Indicates the current elapsed time in dotted lines.

l

Contents of display

The scales are displayed as follows:

Scales/Item

Description

Horizontal Scale

The analysis cycle set for each stream is regarded as a full scale. In Manual

mode, when you issue the measurement start command from the window,

chromatogram display starts and continues until the measurement end

command is issued.

During this period, each time a chromatogram exceeds the horizontal axis,

the time scale is automatically extended and adjusted so that the entire

chromatogram is displayed.

Vertical Scale

The vertical scale is displayed in auto scale and is automatically adjusted so

that data deflection up to that time is about 80% of the range. Auto scaling is

independently achieved for both chromatograms and temperature.

Stream under

Measurement

Indicates the number of the stream currently being measured.

In conditions where sample replacement is possible (calibration

(verification) method can be done in manual only), “Replacement available”

is displayed.

Stream to be Measured

Next

Indicates the number of the stream scheduled to be measured next.

If the stream concerned is under preparation, a “P” indication is added.

n

Timing Display Section

The timing display section is the part of the Analyzer Operation window that displays the ON/OFF

status of the valves for each analysis cycle and the ON/OFF timing of the gate(s).
The current elapsed time is indicated by dotted lines; which timing and where the current location

is in the analysis cycle is recognizable at a glance.
If there are two detectors, the gate ON/OFF timing is displayed on a detector basis.

5th Edition : Nov. 16, 2011-00