5 changing operation mode, Changing operation mode -8 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 55

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-8

IM 11B06C01-01E

4.2.5 Changing Operation Mode

You can change the GC8000 operation mode.

To change the operation mode of the entire GCM, call the Change All GCM Operation Mode

dialog box in either of the following ways.

• Select the Change All GCM Operation Mode command from the Operation Mode menu.
• Click the Change All GCM Operation Mode button on the toolbar.

Figure 4.9

Change All GCM Operation Mode dialog box

From this dialog box (figure 4.9), you can select the Run, Pause, Stop, or Forced Stop mode.

(Note that you can select the Forced Stop if you are at user level C+.)

To change the operation mode of each GCM, call the Change Each GCM Operation Mode dialog

box in either of the following ways.

• From the Operation Mode menu, select the Change Operation Mode command.
• Click the Change Operation Mode button in each GCM display area.

Figure 4.10

Change Each GCM Operation Mode dialog box

From this dialog box (figure 4.10), you can select the Run, Pause, or Stop mode.
In both dialog boxes, you can change the operation mode as follows.
1) Select an operation mode you wish to change, and click the OK button. A confirmation

message showing “The operation mode is changed to xx” appears (where, “xx” is the

selected operation mode). If the GCM is not in the Process status, the “The request for

change of GCMx operation mode was rejected” message is returned (where, “xx” is a GCM

number).

5th Edition : Nov. 16, 2011-00