<Appendix B List of Messages>

B-3

IM 11B06C01-01E

n

Analyzer Operation

Message

Meaning

Countermeasures

[O

pe

ra

tio

n

M

od

e]

Operation command was not

accepted by the analyzer.

An unauthorized level user is

attempting to change the operation

mode or operate a valve.

Only level C users can change the

operation mode or operate valves.

Valves can only be operated

in Manual mode.

An attempt is being made to

operate a valve in a mode other

than the Manual mode.

To operate a valve, choose the

Manual mode. Note that only

level C users can carry out this

operation.

Detectors can be operated

only in Manual mode.

An attempt is being made to

operate a detector in a mode other

than the Manual mode.

To operate a detector, choose

the Manual mode. Note that only

level C users can carry out this

operation.

Operation pattern, valve, or

detector cannot be operated

because the user level is

inappropriate.

An unauthorized level user is

attempting to change the operation

pattern or operate a valve.

Only level B or C users can

change the operation pattern.

Valves can only be operated by

level C users.

[S

et

U

p]

The entered data is out of

range.

An invalid value has been entered

on Valve/Peak setting dialog box.

Enter the correct value.

The entered data is the same

as the current data.

“Send” button has been clicked

when the data is not changed on

the Valve/Peak setting dialog box.

Click the “Send” button after

changing the data.

Sending data to analyzer

failed.

The data transmission has failed

on the Valve/Peak setting dialog

box.

Try again after checking the

operation status of PCAS and

Analyzers, and connection status

of the cables.

Please change the LCD mode

on the GC1000 to remote

mode.

The Valve/Peak setting is not

available when the analyzer is

under “local mode”.

Try again after changing to

“remote mode” at the analyzer.

A user has changed

a parameter in the

GC1000.¥nPlease restart

Analyzer Operation Window

by the [Display][Analyzer

ID] command of Overview

window.

Due to the parameter change by

LCD panel at the analyzer, there

is a descrenpacy of the setting

parameter information on ASET.

The Valve/Peak setting from ASET

is not available.

Re-load the initial database

from the analyzer, then setting is

available.

n

Chromatogram

Message

Meaning

Countermeasures

[T

im

e/

Pr

es

su

re

D

is

pl

ay

]

Please check Temp. or Press.

box.

Temperature/Pressure is not

selected on the Temp/Pressure

dialog box.

Select at least Temperature or

Pressure.

Pe

n

se

le

ct

Select two graphs.

2 graphs are not selected at

Differential chromatogram display.

Select 2 graphs.

Illegal file type

An attempt was made to open a

file other than a chromatogram file.

Select the correct file.

Already opened

The selected differential

chromatogram has already been

opened by another pen.

Select another differential

chromatogram or erase the

opened differential chromatogram.

n

Alarm

Message

Meaning

Countermeasures

H

is

to

ry

er

as

in

g Alarm history cannot be reset

by a level-A user.

An unauthorized level user is

attempting to reset the alarm

history.

Only level B or C users can reset

the alarm history.

5th Edition : Nov. 16, 2011-00