<5. Display and Operations while Connected to GC1000 Mark II>

5-9

IM 11B06C01-01E

l

Contents of display

The following table lists the contents of the display:

Display Item

Description

Stream Valve

Indicates the stream-switching valve status by number. If all the valves are

OFF, this section is grayed out.

V1 to 8

Indicates the ON/OFF status of valves 1 to 8.

Det1/Det2

Indicates the ON/OFF status of the detectors. When the status is in Manual

mode, moving the mouse cursor over this section and clicking the right

button causes the Detector Signal Display window to appear.

HEAT/PRESS/AI

Displays one of the following: temperature controller status, EPC status, or

AI value. Which value to display can be selected using the List Button box.

When the status is in Manual mode and either HEAT or PRESS is

displayed, moving the mouse cursor over this section and clicking the left

button causes the Temperature Control Unit Operation window or Pressure

window to appear respectively.

SEE ALSO

For detector signals and temperature controller status display, see section 3.3.

n

Data Display Section

The data display section is the part of the Analyzer Operation window that displays the elapsed

time, peak number, peak name, and detector signal. The data updating cycle is 1 second. If there

are two detectors, data is displayed on a detector basis.

l

Window elements

F5-2-7.ai

Common data section

Detector 1

data

Detector 2

data

l

Contents of display

The following table lists the contents of the display:

Display Item

Description

Elapsed Time

Indicates the time (second) that has elapsed from the start of analysis.

This section also shows the ratio of the elapsed time to the analysis cycle

in %. In the Manual mode, because the time axis is automatically adjusted,

the ratio of the elapsed time to the full scale is indicated in %. In cases of

being prepared, this part indicates the elapsed time in minus value.

Peak Number

Indicates the number of the peak currently being detected.

Peak Name

Displays the name of the peak currently being detected. Moving the mouse

cursor over the peak name causes the peak name to appear as a tool hint.

Detector Output

Displays a detector signal in mV. The chromatogram displayed in the

chromatogram display section represents the pattern of this signal

5th Edition : Nov. 16, 2011-00