7 gate mark, peak mark, and start mark, Gate mark, peak mark, and start mark -36 – Yokogawa GC8000 Process Gas Chromatograph User Manual

Page 83

Advertising
background image

<4. Display and Operations while Connected to GC8000>

4-36

IM 11B06C01-01E

Figure 4.40

Example of display in auto gain

4.4.7

Gate Mark, Peak Mark, and Start Mark

In mark settings, if you select gate mark, peak mark, and start mark to be displayed, they are

displayed in chromatogram in the enlargement display area.

Gate OFF

Peak OFF

Start mark

Gate ON

Peak ON

Figure 4.41

Example of displaying gate mark, peak mark, and start mark

4.4.8

Additional Information (Peak No., Peak Name, etc.)

Among the following data items, up to two items can be displayed in the Enlargement Display

window.

- Relative peak number, - Peak name (up to 8 characters), - Gain (multiplying factor is

displayed in “G:n” where n = 0 to 15), - Range (displayed in “Rng: XX.XX unit”), - Standard

concentration (displayed in “S.Cnc:XX.XX unit”).

Click on the Set Mark command in the Graph menu to display the Set Mark dialog box. Select

data from the additional information 1/additional information 2 lists. If you want to hide data,

select “None”. In case of off-line, peak number is only indicated.

The following shows examples of displaying additional information.

Range and

standard concentration

Peak name

and gain

Figure 4.42

Examples of displaying gate mark and peak mark

5th Edition : Nov. 16, 2011-00