<Appendix B List of Messages>

B-5

IM 11B06C01-01E

Message

Meaning

Countermeasures

G

ra

ph

p

lo

tti

ng

Graph cannot be plotted.

Because of no data

A graph cannot be plotted because

the currently displayed window

contains no data.

Perform graph plotting only when

data is available in the window.

Graph cannot be plotted.

Illegal data

No graph was plotted because

the graph range to plot contained

inappropriate data.

This message also appears if

there is no data in the graph range.

Check if data in the set graph

range is appropriate.

Also check if there is data in the

graph range.

Selection error

Data required is not

selected.

Some of the data required to plot

the graph have not been selected.

Check that the complete data

required to plot the graph has

been provided.

Selection error

The maximum number of

peaks that can be selected

at one time is 6.

More than 6 peaks have been

selected.

The maximum number of peaks

that can be displayed on a graph

at one time is 6.

Select six or less peaks.

Selection error

Start time was set to a time

later than the end time.

The start time was set to a time

later than the end time when

setting up the time axis (in Manual

mode).

Set the start time to a time earlier

than the end time.

Graph cannot be created /

plotted.

For logarithmic graphs,

axes can not be set at zero

or negative values.

Zero or negative values were

set as the minimum or maximum

values when setting up logarithmic

axes (in Manual mode).

Logarithmic axes cannot display

zero or negative values. When

setting the maximum and

minimum values for the logarithmic

axes, set values greater than zero.

n

ASIU

Message

Meaning

Countermeasures

[D

O

O

pe

ra

tio

n] ASIU operation is not

available in User level A.

An attempt is being made to

operate ON/OFF of DOs by an

inappropriate user level.

User level B or C can operate the

ON/OFF of DOs.

n

File extensions

In each file reference window, only files having the following extensions can be opened:

File analysis results:

.ana (latest analysis results)

.nar (latest analysis results)

File concentration analysis history: .cnc (concentration analysis history)

.nar (latest analysis results)

.dat (ASIU average value)

File Retention time history:

.rtm (Retention time history)

.nar (latest analysis results)

File laboratory analysis results:

.lab (laboratory analysis results)

.dat/old (laboratory analysis results)

File calibration coefficient history: .cef (calibration coefficient history)

.dat/old (calibration coefficient history)