<4. Display and Operations while Connected to GC8000>

4-15

IM 11B06C01-01E

Select a Calibration/Validation Method of either Manual, Semi-automatic or Auto, and click the

Execute button. The Calibration/Validation Method is changed. The current Calibration/Validation

Method is selected in the dialog box if it is opened.
If the initial analyzer database has been updated when you call this dialog box, the following

message is displayed and you cannot start this dialog box. In such case, reload the initial

database and recall this dialog box.

Figure 4.19

The message displayed if the initial analyzer database has been updated

4.3.8

Changing Auto Calibration Status

To change the Auto Calibration Status, select the Auto Calibration Status command from the

Measurement Status menu, and click the Start button. The Auto Calibration function of active

GCM is enabled. When you select “Stop” from the Auto Calibration Status command, the Auto

Calibration function is stopped. You can also change the Auto Calibration Status of active GCM

using the “Start/Stop Auto Calibration” button on the toolbar.
On the Start/Stop submenu, the currently selected option is checked.
If the Calibration/Validation Method is set to Semi-automatic or Manual, the Auto Calibration

Status command is grayed out on the Measurement Status menu and it is inoperable. Also, the

Start/Stop Auto Calibration button is grayed out on the toolbar.

4.3.9

Operating Stream Valve

To operate a stream valve, set the GCM to the Manual status and select the Manual or Stream

VLV command from the Manual/Range menu. The Operating Stream Valve dialog box appears.

Figure 4.20

Operating Stream Valve dialog box

To turn the stream valve On, select ON. The stream number list of the GCM is displayed from the

ON Stream Valve Number combo box. Then, click the Execute button.
To turn the stream valve Off, select OFF and click the Execute button.
You can also call this Operating Stream Valve dialog box by clicking the stream number display of

the Analyzer Operation window.