<5. Display and Operations while Connected to GC1000 Mark II>

5-18

IM 11B06C01-01E

A dialog box appears confirming that you want to change the ON/OFF setting.

F5-2-21.ai

(3) Click the Yes button

The relevant State/Remark entry changes to ON or OFF.

5.2.8

Changing Range

The range of each peak can be changed using the range table that has been set to the analyzer.
This section describes how to change the range.

CAUTION

Only level B or C users can change the range.
The analyzer must also be in Remote mode.

n

Changing Range

(1) Click the Range Change Command button

.

(2) The Range Change dialog box appears.

F5-2-22.ai

Specifying the stream number, peak number, and range number you wish to change causes

the current information of each item to appear.

(3) Click the Execute button. This causes a confirmation message to appear.