App index – Yokogawa Standalone MW100 User Manual

Page 205

Advertising
background image

4-13

IM MW100-01E

Troubleshooting and Maintenance

1

2

3

4

5

App

Index

• When calibrating the RTD range of the 4-CH, High-Speed Universal Input module

100Vpk MAX TO

250V MAX CH TO CH

600V MAX TO

100Vpk MAX TO

250V MAX CH TO CH

600V MAX TO

H
L
G

Resistance standard

Input terminal

Input terminal

Short

Make the resistance of

three lead wires equal.

When calibrating 0 Ω

When calibrating a range other than 0 Ω

* Carry out calibration for each input

terminal.

b

A

B

b

A

B

• When calibrating the DC voltage range of the 10-CH, Medium-Speed Universal Input

module

100Vpk MAX TO

120V MAX CH TO CH

600V MAX TO

Hi

Lo

Input terminal of CH2

Short the input terminal of CH1 (apply 0 V)

+

+

DC voltage/current standard

• When calibrating the RTD range of the 10-CH, Medium-Speed Universal Input module

b

A

B

b

A

B

100Vpk MAX TO

120V MAX CH TO CH

600V MAX TO

H
L
G

Short the input terminal of CH3 (connect 0 Ω)

When calibrating the RTD (1 mA) 600 mV range,

connect 300 Ω to the input terminal of CH5.*

Make the resistance of three lead wires equal.

* When calibrating the RTD (1 mA) 60 mV range, connect 60

Ω

to

the input terminal of CH6.

When calibrating the RTD (1 mA) 200 mV range, connect 200

Ω

to the input terminal of CH4.

Resistance standard

4.4 Calibration

Advertising