Operation

9

2.4.1

APC4000/APC4001 Interface Option Enable

The APC4000 interface enable is activated only for models that are purchased and used with the PCM1000-1
controller or part of an APC4001 controller system. When UPS3000 or UPS3210 is used in conjunction with a
controller, Peak Hold, Freeze Mode or MIN/MAX Mode options are not available.

NOTE: UPS3110 is not used with a controller.

Approximately 1994 multi conversion software was added to UPS3000 models, therefore UPS3000 and PCM1000 units
made prior to this will not work with the ones manufactured after that date. Consult factory for information on upgrading
units.

2.4.2

Peak Hold or MAX/MIN Option Enable

See “Peak Hold Option” on page 54 or “Min and Max Mode Option” on page 55.

NOTE: * Absolute or Gage modes of operation are factory set, depending upon the style of transducer supplied and the
type of internal memory configuration utilized.

When the Absolute/Gage switching is not utilized; the front panel

ABS/GAGE

push-button switch will be

programmed to inactive and covered with a blank overlay.

When the PEAK HOLD Option or MAX/MIN option is supplied switching from Absolute to Gauge mode via
the front panel is possible and the Freeze Mode option, if supplied, is operable.

NOTE: When the PEAK HOLD or MAX/MIN option is supplied, the CONV button is no longer active or present on front
panel. Therefore, the required conversion must be set thru reconfiguration. See “Engineering Conversion with PEAK
HOLD or MAX/MIN Option” on page 12 for appli
cable method to change to required conversion.

Switch Settings S1

Position 1

Position 2

Position 3

Position 4

Position 5

Position 6

Position 7

Position 8

UPS 3000
UPS 3110
UPS 3210

0

0

1

0

0

0

1

1

With Controller

0

0

1

0

0

0

1

0

Without

Controller

Table 2-1. APC 4000/APC 4001 Interface Enable Switch Settings

Switch Settings S1

Peak Hold or MAX/MIN Mode

*ABS/GAGE Mode

Position 1

Position 2

0

0

Disabled

ABS & Gauge (Front Panel Switch Selectable)

0

1

Disabled

Gauge or ABS only

1

0

Enabled

ABS & Gauge (Front Panel Switch Selectable)

1

0

Enabled

Gauge or ABS only

Table 2-2. Peak Hold or MAX/MIN Enable Switch Settings