Calibration and Adjustment Procedure

27

value of the Negative Delta Pressure, -0.05 PSI as in this example.

6. Depress the

ENTER

push-button on the UPS3110. The display reads zero. Without touching the

VERNIER

(11) knob repeat this step for the middle and high ranges.

7. Using the

STEPPER

push-button, as shown in Figure 3-2 on page 20, place the unit into

SHUNT RESISTOR

CALIBRATION

mode. The display will be as shown in Figure 3-5 on page 23.

8. With UPS3110

RANGE SELECT

switch (6) switched in the high range, depress the

ZERO

push-button on

the front panel. Verify that the display reads 0.00. Upon release of the push-button the display will be as
shown in Figure 3-5 on page 23.

9. Depress the

ENTER

push-button. The display will respond with

“[] [] [] [] [] []”

until the button is

released.

10. Pressing the

STEPPER

push-button bring the indicator back to its initial Data Recall display condition as

shown in Figure 3-2 on page 20. The display shows

1 XXX

(three arbitrary digits).

11. Depress the

DATA ENTER

switch, S2, located approximately in the middle of the circuit board. See Figure

2-7 for the switch location

12. If the data is accepted, the three-digit number on the right side of the display will indicate

377

for as long

as the

DATA ENTER

switch, S2, is depressed.

13. The barometric offset is now complete and

CALIBRATE/OPERATE

switch, S3, must be returned to its

normal operating positions as shown in Table 3-1 on page 20.