22

UPS3000/UPS3110/UPS3210 Operation and Maintenance Manual

2. Perform the action as indicated when the readings are stable.

When step number 11 is reached, the display will change so that the left most status symbol will be “H”. This
will remain for step 12 and down to approximately 0.00 psi.

NOTES:

1. If reading is in motion or correction required is not within ±0.8% of Full-Scale, no entry will be made.

2. If entry is valid, the display will momentarily indicate the correction value (in percent) and the memory location at
which it is stored.

3. If 100% ±0.05% is not obtained, repeat the Zero/Span calibration sequence.

STEP NO.

INPUT PRESSURE,

% OF RANGE

OPERATOR ACTION

REQUIRED

STATUS SYMBOL IN

LEFT MOST DIGIT

REMARKS

1

0

Press ZERO Switch

“Upper Circle”

Zero on Display

2

10

Press ENTER Switch

Notes 1 & 2

3

20

Notes 1 & 2

4

30

Notes 1 & 2

5

40

Notes 1 & 2

6

50

Notes 1 & 2

7

60

Notes 1 & 2

8

70

Notes 1 & 2

9

80

Notes 1 & 2

10

90

Notes 1 & 2

11

100

No Action Required

“Lower Circle”

Note 3

12

50

Press ENTER Switch

Notes 1 & 2

13

0

No Action Required

“L”

Table 3-3. LINEARIZATION & HYSTERESIS CALIBRATION SEQUENCE