Changing the battery, Rt‐10 flow monitor, Battery replacement – AW Gear Meters RT-10 User Manual

Page 22

Advertising
background image

RT‐10 Flow Monitor

 

Operation and Programming Manual 

 

 

 

21 

      
 

Changing the Battery 

Remove the enclosure cover.  

 

You must remove the cover outside the hazardous area if dangerous conditions are present. 

When the battery timer has expired, LO BAT displays briefly every time you change the display mode to 
indicate that the expected battery life has been reached. The battery is a 3.6 volt Lithium “C” cell 
available from AW Gear Meters (part number MI50004).  

 

 

 

 

 

 

Battery Replacement 

To change the battery, follow these steps: 

1. Locate the four screws recessed behind the faceplate.  
2. Back these screws out (screws are retained and do not come completely out).  
3. Carefully remove the faceplate/circuit board assembly and turn it over.  
4. A tie‐wrap secures the battery in the battery holder. Cut the tie‐wrap. 
5. Install new battery making certain to install it in the proper orientation (positive [+] end to the 

contact with the red wire, the negative [‐] end to the contact with the black wire).  

6. Immediately turn over the assembly over and follow the procedure for resetting the battery 

timer (below).  

7. When finished with reset procedure, install new tie‐wrap around battery.  
8. Reinstall faceplate assembly into housing. 

 

 

SR# RT001500

UP

SEL

ENT

RT-EX10

UP

SEL

ENT

DN/

MODE

DN/

MODE

LO O SEN
SC REW S

B A C K- O UT FO UR SC REW S
PULL O UT A SSEM B LY
(SC REW S A RE RETA IN ED )

TURN O VER A SSEM B LY

REM O VE TIE W RA P

C HA N GE B A TTERY

IN STA LL N EW TIE W RA P

LO O SEN
SC REW S

- TO B LA C K W IRE

+ TO RED W IRE

O B SERVE PO LA RITY