Rt‐10 flow monitor – AW Gear Meters RT-10 User Manual
Page 23
RT‐10 Flow Monitor
Operation and Programming Manual
22
Resetting the Battery Timer
As soon as you place a battery in the holder, the RT‐10 alternately displays “bA old” (battery old) and
“bA neu” (battery new) several times for approximately one second each. If you have replaced the
battery with a new battery, press ENT when “bA neu” appears to reset the battery timer to zero days.
If you re‐installed the same battery, press ENT while “bA old” appears; the battery timer will not reset.
If you do not press ENT before this procedure exits, the timer will not reset.
To return to the reset procedure, briefly remove battery again.
If you do not peform a battery timer reset, the RT‐10 continues to display LO BAT briefly whenever the
mode changes.
The device maintains total values in memory by battery power. These total values are backed‐up in non‐
volatile memory every 24 hours when the clock reaches 12:00 A.M. The total value accumulated since
midnight is lost if you change or remove the battery.
1 SEC
1 SEC
EN T
EN T
SA M E
B A TTERY
N EW
B A TTERY
O N B A TTERY C HA N GE
IF B A TT. IS REM O VED /REPLA C ED D ISPLA Y W ILL B LIN K O LD /N EW .
PRESS EN T W HILE N EW IS VISIB LE TO RESET B A TT. TIM ER FO R N EW B A TTERY.
PRESS EN T W HILE O LD IS VISIB LE IF B A TTERY W A S N O T C HA N GED .
Notice
Notice