Field Installation

Connecting and Deploying Instruments

Ensure data-logging capable instruments have been programmed with a data log
before completing the following steps. Refer to the instrument manual for instructions on
programming a data log. Ensure alarm conditions have been programmed (if
applicable).

Ensure all instruments have been properly, uniquely addressed in Win-Situ Software
with the correct Modbus settings.

Perform the following steps for each instrument. Ensure you are installing the
instrument in its proper location.

1. Connect the instrument to the RuggedCable using the twist-lock connector. The

twist-lock connector clicks when the connection is tight.

2. Remove the red dust cap from the bulkhead connector of the enclosure.

3. Connect the up-hole end of the cable to the bulkhead connector of the enclosure.

4. Install the instrument in its deployment location. Refer to the instrument Operator's

Manual for any questions regarding proper connection or deployment.

5. Ensure the cable does not apply excessive weight on the enclosure by securing the

cable grip.

Installing the TROLL Link Telemetry System

To mount the enclosure, you may obtain a mounting tripod or install your own pole.
Install the pole before proceeding.

1. Hold the enclosure to the pole at the desired height (be sure to leave room at the top

of the pole for the solar panel).

2. Place a hose clamp or U-bolt into the two cutouts of the top bracket of the enclosure

and around the pole.

3. For hose clamps tighten the clamp until the enclosure does not move. Attach the

second hose clamp to the bottom of the enclosure.
For U-bolts Place the smaller bracket over the open ends of the bolt. Use split-lock
washers and nuts to tighten the assembly to the pole.
Repeat for any remaining open brackets.

4. Remove the two yellow caps from the vent tube inside the enclosure. Attach the

desiccant tube to the vent tube.