SECTION 5 – CALIBrATION OF THE MICrO VOLuME

ACCESSOry

Jenway recommends that the micro volume accessory is calibrated every 6 months. A set of calibration
solutions (Part code 035 092) are supplied with the Genova Nano spectrophotometer.

5.1

CALIBrATION SOLuTIONS (035 092)

The supplied calibration solution set consists of two
vials.

1. Matrix Blank (White)

2. 10x ref - Calibration Standard (Blue)

A calibration certificate is supplied that details the
certified absorbance values of the calibration standard
and the path length at which these values were
determined.

5.2

ACCESSOry CALIBrATION PrOCEdurE

5.2.1

Micro Volume Accessory Settings

The micro volume accessory settings can be accessed
by pressing the key below the Micro Volume Accessory
icon in any of the Genova Nano’s operating modes.

5.2.2

Calibration Standard - data Entry

1. Enter the A260 and A330 DNA standard solution
values from the supplied calibration certificate
into the accessory settings by pressing the key
adjacent to the Calibration Standard Value icon.

16

SECTION 5 – CALIBRATION OF THE MICRO VOLUME ACCESSORY

Jenway recommends that the micro volume accessory is calibrated every 6 months. A set of
calibration solutions (Part code 035 092) are supplied with the Genova Nano spectrophotometer.

5.1

CALIBRATION SOLUTIONS (035 092)

The supplied calibration solution set consists of
two vials.

1. Matrix Blank (White)

2. 10x ref - Calibration Standard (Blue)

A calibration certificate is supplied that details the
certified absorbance values of the calibration
standard and the path length at which these values
were determined.

5.2

ACCESSORY CALIBRATION PROCEDURE

5.2.1

Micro Volume Accessory Settings

The micro volume accessory settings can be
accessed by pressing the key below the Micro
Volume Accessory icon in any of the Genova
Nano’s operating modes.

5.2.2

Calibration Standard - Data Entry

1. Enter the A260 and A330 DNA standard

solution values from the supplied calibration
certificate into the accessory settings by
pressing the key adjacent to the Calibration
Standard Value icon.

16

SECTION 5 – CALIBRATION OF THE MICRO VOLUME ACCESSORY

Jenway recommends that the micro volume accessory is calibrated every 6 months. A set of
calibration solutions (Part code 035 092) are supplied with the Genova Nano spectrophotometer.

5.1

CALIBRATION SOLUTIONS (035 092)

The supplied calibration solution set consists of
two vials.

1. Matrix Blank (White)

2. 10x ref - Calibration Standard (Blue)

A calibration certificate is supplied that details the
certified absorbance values of the calibration
standard and the path length at which these values
were determined.

5.2

ACCESSORY CALIBRATION PROCEDURE

5.2.1

Micro Volume Accessory Settings

The micro volume accessory settings can be
accessed by pressing the key below the Micro
Volume Accessory icon in any of the Genova
Nano’s operating modes.

5.2.2

Calibration Standard - Data Entry

1. Enter the A260 and A330 DNA standard

solution values from the supplied calibration
certificate into the accessory settings by
pressing the key adjacent to the Calibration
Standard Value icon.

Micro Volume Accessory icon

1

2

Blue
label

White
label