Jenway Genova Nano English User Manual

Page 15

Advertising
background image

13

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys adjacent to
the arrow icons to increase or decrease the number. Press the key adjacent to the tick icon to save
any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

17

2. Select the digit to be changed using the keys

at the bottom of the screen. Use the keys
adjacent to the arrow icons to increase or
decrease the number. Press the key adjacent
to the tick icon to save any changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for
the DNA standard solution values, in mm, into
the accessory settings by pressing the key
adjacent to the Calibration Standard Path
Length Value icon.

4. Select the digit to be changed using the keys at the bottom of the screen. Use the keys

adjacent to the arrow icons to increase or decrease the number. Press the key adjacent to the
tick icon to save any changes.

5.2.3

Accessory Calibration and Verification

If an error message is displayed during the accessory calibration and verification procedure,
refer to Section 9 for further details.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path
Length Calibration icon.

2. The Air Measurement icon will be displayed.

Use a lint free cloth to clean the upper and
lower read heads, close the instrument lid and
press the key adjacent to the tick icon to
continue. Pressing the key adjacent to the
cross icon will return the instrument to the
micro volume accessory settings menu screen.

2. Select the digit to be changed using the keys at

the bottom of the screen. Use the keys adjacent
to the arrow icons to increase or decrease the
number.

Press the key adjacent to the tick icon to save any
changes.

3. Enter the stated path length (1.0mm for the

supplied calibration standard solution set) for the
DNA standard solution values, in mm, into the
accessory settings by pressing the key adjacent
to the Calibration Standard Path Length Value
icon.

1. The accessory calibration sequence is initiated

by pressing the key adjacent to the Path Length
Calibration icon.

2. The Air Measurement icon will be displayed. Use

a lint free cloth to clean the upper and lower read
heads, close the instrument lid and press the key
adjacent to the tick icon to continue. Pressing
the key adjacent to the cross icon will return
the instrument to the micro volume accessory
settings menu screen.

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.