5

SECTION 2 – ACCESSOry LAyOuT ANd INSTALLATION

2.1

ACCESSOry LAyOuT






Fig 2.1 – Micro Volume Accessory Layout

2.2

ACCESSOry rEMOVAL

The micro volume accessory will be pre-installed when purchased as part of the Genova Nano
spectrophotometer. The accessory is optically matched to the spectrophotometer in which it is originally
installed. Users are therefore not able to transfer the micro volume accessory to other Genova Plus/Nano
spectrophotometers without the assistance of trained service personnel.

The micro volume accessory should not be removed or re-installed un-necessarily as this may
affect the accuracy of the accessories optical alignment.

The following steps must be followed to remove and re-install the micro volume accessory into the
Genova Nano spectrophotometer.

Remove the front panel from the sample chamber
by loosening screws 1 and 2 until the front panel
can be lifted out in the forward direction.

Linkage Connection

Upper Read Head

Read Head Cover

Path Length
Drive Motor

Side Covers

Lower Read Head

Remove front panel

1

2