SECTION 9 – TrOuBLESHOOTING

9.1

ACCESSOry ErrOr COdES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete calibra-
tion process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of acceptable range

2

Blank solution reading, 0.2mm path length, outside of acceptable range

3

Calibration solution reading, 0.5mm path length, outside of acceptable

range

4

Calibration solution reading, 0.2mm path length, outside of acceptable

range

5

Duplicate calibration solution readings, 0.5mm path length, outside of

acceptable range

6

Duplicate calibration solution readings, 0.2mm path length, outside of

acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of acceptable range

10

Blank solution reading, 0.2mm path length, outside of acceptable range

11

Calibration solution reading, 0.5mm path length, outside of acceptable

range

12

Calibration solution reading, 0.2mm path length, outside of acceptable

range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

23

SECTION 9 – TROUBLESHOOTING

9.1

ACCESSORY ERROR CODES

If an error code is displayed it will be accompanied by a clipboard icon and a number to indicate the
cause of the error. The first time that an error is displayed, it will be possible to repeat the erroneous
measurement. If the same error message is shown after the repeat measurement, the complete
calibration process will need to be repeated. The table below shows the error codes:

9.1.1

Calibration Procedure Error Codes

Error Code

Symbol

Issue

1

Blank solution reading, 0.5mm path length, outside of

acceptable range

2

Blank solution reading, 0.2mm path length, outside of

acceptable range

3

Calibration solution reading, 0.5mm path length, outside of

acceptable range

4

Calibration solution reading, 0.2mm path length, outside of

acceptable range

5

Duplicate calibration solution readings, 0.5mm path length,

outside of acceptable range

6

Duplicate calibration solution readings, 0.2mm path length,

outside of acceptable range

7

0.2mm path length factor outside of acceptable range

8

0.5mm path length factor outside of acceptable range

9.1.2

Verification Check Error Codes

9

Blank solution reading, 0.5mm path length, outside of

acceptable range

10

Blank solution reading, 0.2mm path length, outside of

acceptable range

11

Calibration solution reading, 0.5mm path length, outside of

acceptable range

12

Calibration solution reading, 0.2mm path length, outside of

acceptable range

Advertising