Jenway Genova Nano English User Manual

Page 16

Advertising
background image

14

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

18

The unit takes five dark and light readings at
260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

5. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

6. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

9. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

10. The Calibration Solution Measurement icon will

be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration solution

onto the lower read head. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the calibration solution measurement.

3. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution
onto the lower read head.

Close the instrument lid and press the key
adjacent to the tick icon to initiate the blank
measurement.

4. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

6. The Calibration Solution Measurement icon

will be displayed. Open the instrument lid and
use a lint free cloth to clean the read heads.

7. Pipette a 2.0μl aliquot of the calibration

solution onto the lower read head. Close the
instrument lid and press the key adjacent to
the tick icon to initiate the calibration solution
measurement.

8. The unit takes five readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

10. The Calibration Solution Measurement icon

will be displayed for a second time. Open the
instrument lid and use a lint free cloth to clean
the read heads.

11. Pipette a 2.0μl aliquot of the calibration

solution onto the lower read head. Close the
instrument lid and press the key adjacent to
the tick icon to initiate the calibration solution
measurement.

5. If the measured values are within the required tolerances, the Passes Test icon will be displayed. Press

the button adjacent to the Tick icon to continue or the button adjacent to the Cross icon to cancel
the calibration process.

9. If the measured values are within the required tolerances, the Passes Test icon will be displayed. Press

the button adjacent to the Tick icon to continue or the button adjacent to the Cross icon to cancel
the calibration process.

12. The unit takes five readings at 260 and 330nm, at both the 0.5 and 0.2mm path lengths.

13. If the measured values are within the required tolerances, the Passes Test icon will be displayed and

the instrument will calculate the new path length calibration factors. Press the button adjacent
to the Tick icon to continue or the button adjacent to the Cross icon to cancel the calibration
process.

14. The new calibration factor values must now be verified by re-measuring the air, blank and standard

values.

Advertising