12. The unit takes five readings at 260 and 330nm, at both the 0.5 and 0.2mm path lengths.

13. If the measured values are within the required tolerances, the Passes Test icon will be

displayed and the instrument will calculate the new path length calibration factors. Press the
button adjacent to the Tick icon to continue or the button adjacent to the Cross icon to cancel
the calibration process.

14. The new calibration factor values must now be verified by re-measuring the air, blank and

standard values.

15. The Air Measurement icon will be displayed.

Open the instrument lid and use a lint free
cloth to clean the read heads. Close the
instrument lid and press the key adjacent to
the tick icon to initiate the blank
measurement.

16. The unit takes three dark and light readings at

260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

17. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution
onto the lower read head. Close the
instrument lid and press the key adjacent to
the tick icon to initiate the blank
measurement.

18. The unit takes three readings at 260 and

330nm, at both the 0.5 and 0.2mm
pathlengths.

19. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

19

12. The unit takes five readings at 260 and 330nm, at both the 0.5 and 0.2mm path lengths.

13. If the measured values are within the required tolerances, the Passes Test icon will be

displayed and the instrument will calculate the new path length calibration factors. Press the
button adjacent to the Tick icon to continue or the button adjacent to the Cross icon to cancel
the calibration process.

14. The new calibration factor values must now be verified by re-measuring the air, blank and

standard values.

15. The Air Measurement icon will be displayed.

Open the instrument lid and use a lint free
cloth to clean the read heads. Close the
instrument lid and press the key adjacent to
the tick icon to initiate the blank
measurement.

16. The unit takes three dark and light readings at

260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

17. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution
onto the lower read head. Close the
instrument lid and press the key adjacent to
the tick icon to initiate the blank
measurement.

18. The unit takes three readings at 260 and

330nm, at both the 0.5 and 0.2mm
pathlengths.

19. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process.

15. The Air Measurement icon will be displayed.

Open the instrument lid and use a lint free cloth
to clean the read heads. Close the instrument
lid and press the key adjacent to the tick icon to
initiate the blank measurement.

16. The unit takes three dark and light readings at

260 and 330nm, at both the 0.5 and 0.2mm
pathlengths.

17. The Blank Solution Measurement icon will be

displayed. Open the instrument lid and pipette
a 2.0μl aliquot of the matrix blank solution onto
the lower read head. Close the instrument lid
and press the key adjacent to the tick icon to
initiate the blank measurement.

18. The unit takes three readings at 260 and 330nm,

at both the 0.5 and 0.2mm pathlengths.

19. If the measured values are within the required tolerances, the Passes Test icon will be displayed.

Press the button adjacent to the Tick icon to continue or the button adjacent to the Cross icon to
cancel the calibration process.

21. The unit takes three readings at 260 and 330nm, at both the 0.5 and 0.2mm pathlengths.

22. If the measured values are within the required tolerances, the Passes Test icon will be displayed.

Press the button adjacent to the Tick icon to continue or the button adjacent to the Cross icon to
cancel the calibration process.

20

20. The Calibration Solution Measurement icon

will be displayed. Open the instrument lid and
use a lint free cloth to clean the read heads
Open the instrument lid and pipette a 2.0μl
aliquot of the calibration solution onto the
lower read head. Close the instrument lid and
press the key adjacent to the tick icon to
initiate the calibration solution measurement.

21. The unit takes three readings at 260 and 330nm, at both the 0.5 and 0.2mm pathlengths.

22. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process

23. The instrument display will return to the micro

volume accessory settings screen and the
new path length calibration factors will be
shown.

20

20. The Calibration Solution Measurement icon

will be displayed. Open the instrument lid and
use a lint free cloth to clean the read heads
Open the instrument lid and pipette a 2.0μl
aliquot of the calibration solution onto the
lower read head. Close the instrument lid and
press the key adjacent to the tick icon to
initiate the calibration solution measurement.

21. The unit takes three readings at 260 and 330nm, at both the 0.5 and 0.2mm pathlengths.

22. If the measured values are within the required tolerances, the Passes Test icon will be

displayed. Press the button adjacent to the Tick icon to continue or the button adjacent to the
Cross icon to cancel the calibration process

23. The instrument display will return to the micro

volume accessory settings screen and the
new path length calibration factors will be
shown.

20. The Calibration Solution Measurement icon will

be displayed. Open the instrument lid and use
a lint free cloth to clean the read heads Open
the instrument lid and pipette a 2.0μl aliquot
of the calibration solution onto the lower read
head. Close the instrument lid and press the key
adjacent to the tick icon to initiate the calibration
solution measurement.

23. The instrument display will return to the micro

volume accessory settings screen and the new
path length calibration factors will be shown.