6. Operation

6.1 Description

Here is described a series of steps from connecting a measuring head to measuring with
the head.

Connect the measuring head or the AE2000 (A/E
converter), and the accessory power cable to the rear
of the instrument.

Turn on the power switch at the rear of the instrument
to supply the power of 85 to 264 VAC (50/60 Hz).

Immediately after the startup, the instrument falls in
the master request condition (“MAS REQ” appears

on the alphanumeric display). Press one of the

keys to clear the master request.

The instrument changes to the measure mode.

In the measure mode, press the

key for 2 sec or more to change to the set mode.

Enter the set value. Refer to “6.4 Setting” and “Set
mode” in “13. Operation flow”.

In the measure mode, press the

key for 2 sec or more, and press the

key once

to change to the adjust detector mode.

Adjust the detector or AE2000. Refer to “Adjust
detector mode” in “13. Operation flow”.

In the adjust detector mode, press the

key twice to change to the master mode

(showing “MIN M.” on the alphanumeric display).

Calibrate the instrument with a master. Refer to “6.3
Automatic mastering” and “Master mode” in “13.
Operation flow”.

In the measure mode, press the

key once to

change to the hold measured value condition.
Press the

key to clear the measured value hold.

The measured value cannot be held when the
mastering result is NG.
Refer to “6.2 Measuring”.

Connecting

1

Turning on power

2

Entering set value

3

Adjusting detector
or AE2000

4

Mastering

5

Measuring

6