11. Troubleshooting

Phenomenon

Cause

Remedy

Repeatability is not stable.

Measuring head is loosely

secured.

Retighten its fasteners.

Specified power is not

supplied.

Supply power of 85 to 264

VAC.

Detector is improperly

adjusted.

Adjust detector.

Instrument falls in hold

measured value
condition. Main display
appears green (dark) or
red (dark).

Press RST key to clear

this condition.

Main display does not
operate.

Instrument is in set mode. ➃ End set mode.

Power is not supplied.

Supply power of 85 to 264

VAC.

Main display and
alphanumeric display do not
light.

Power supply or internal

circuit is faulty.

Ask NIDEC TOSOK for

repair.

12. Cautions

(1)Power cable
The power cable supplied as standard accessory is for 100 V. If you use supply voltage
in excess of 125 V, separately prepare a 250-V power cable.

(2)Control keys
Never operate the control keys with a sharp-pointed tool or the like.

(3)Export
When you try to export this instrument overseas, you may have to have the export
approved by the Ministry of Economy, Trade and Industry under the Export Trade Control
Ordinance. In such a case, contact your nearest NIDEC TOSOK sales office.

(4)Specifications
The specifications are subject to change without notice.