3 automatic measuring functions – SHIMPO DEG-2000 User Manual

Page 17

Advertising
background image

15

[RANK]
The OK ranking method of each item and the number of OK ranks are set.

[LIMIT]
The OK rank boundary of each item is set.

Refer to “Set mode (5)” in “13. Operation flow”.

6.4.3 Automatic measuring functions

The automatic measuring functions are enabled when “+PEAK”, “-PEAK”, “TIR”, and
“TIR/2” are selected with [FUNCTION]. The following automatic measuring functions
are set with [SYSTEM] in the set mode.

[AutoMeas]
Sets the percentage of all master data in the indicating range at which to start the
automatic measurement start function. Disabled when the automatic measurement
start function is turned off.

[InToOut]
Sets whether to hold or reset the measured value when the mastering data change from
inside of the measuring range to outside of the measuring range. Disabled when the
automatic measurement start function is turned off.

[WaitTime]
Sets the stabilizing timer at the time for the automatic measurement to start. When the
stabilizing timer is set at 0.00 sec, the automatic measurement start function is turned off.

[MeasTime]
Sets the measuring timer at the time for the automatic measurement to stop. When the
measuring timer is set at 0.0 sec, the automatic measurement stop function is turned off.

When [RANK] = EQUAL, [RANK n] = 4, [R1/-NG] = -30, and

[+NG/Rn] = +30 are set…

-NG < -30

≤ R1 < -15 ≤ R2 < 0 ≤ R3 ≤ +15 < R4 ≤ +30 < +NG

When [RANK] = SELECT, [RANK n] = 3, [R1/-NG] = -15,

[R1/R2] = -5, [R2/R3] = 0, and [+NG/Rn] = +15 are set…

-NG < -15

≤ R1 < -5 ≤ R2 ≤ 0 < R3 ≤ +15 < +NG

Example

When [WaitTime] = 0.0 and [MeasTime] = 0.0 are set…

Press the ENT key to start the measurement and the RST key to
stop the measurement.

When [I-RANGE (indicating range)] = ±100, [WaitTime] = 1.00,

[AutoMeas] = ±80%, and [MeasTime] = 5.0 are set…

The automatic measurement starts 1 sec after the mastering data
fall within the range of -80 to +80 and stops 5 sec after the start.
The instrument remains in the hold measured value mode until
the measured value falls outside the range of -80 to +80.

Example