7.3.2 Input signal arrangement

Note:

The input signals are enabled when “READY” is turned on. The following
table shows the input signal arrangement of the DC input/output connector
(refer to ◯

15

on page 7).

Pin No.

Signal name

Description

1

Measure command

Hold measured value and data output command
(Enabled when mastering result is OK.)

2

Minimum mastering

Minimum mastering command

3

Maximum mastering

Maximum mastering command

4

RESET

Measured value hold clear command

5

Program change (lower)

6

Program change (higher)

Specifies program numbers 1 to 3 to be changed.
(Refer to Table 6.)

7

Item bit (less significant)

8

Item bit (more significant)

Specifies measurement items 1 to 3 to be output.
(Refer to Table 7.)

9

+COM

10

+COM

Input common line: +12/24 V

Table 6. Program change and program numbers.

Program number

1

2

3

Program change (lower)

On

Off

On

Program change (higher)

Off

On

On

Table 7. Item bits and measurement items to be output.

Measurement item to be output

Setting for this instrument*

1

1

2

3

Item bit (less
significant)

Off

On

Off

On

Item bit (more
significant)

Off

Off

On

On

*

1

Item set with [JUDG OUT] is output.