TC-3000, TC-3000X SPECIFICATIONS

TC-3000

TC-3000X

Temperature

Temperature range

4°C to 99°C

4°C to 99°C

Temperature set point precision

0.1°C

0.1°C

Block uniformity (over full range)

± 0.1°C

± 0.2°C

Temperature accuracy

± 0.2°C

± 0.2°C

End of program cooling below ambient

Yes

Yes

Heating/Cooling Rate

Heating rate, between 55°C and 90°C
on a 50°C to 95°C step:

0.2 ml micro tubes

3.6°C/sec

3.3°C/sec

0.5 ml micro tubes

3.6°C/sec

3.3°C/sec

Cooling rate, between 90°C and 55°C
on a 95°C to 50°C step:

0.2 ml micro tubes

2.0°C/sec

2.0°C/sec

0.5 ml micro tubes

1.8°C/sec

2.0°C/sec

Heated Lid

Adjustable heated lid

No

Yes

Heated lid enable/disable

Yes

Yes

Selectable heated lid temperature

100°C to 115°C

100°C to 115°C

Over-temperature cut-out

Fixed at 145°C

Fixed at 145°C

Temperature sensor

Thermistor

Thermistor

Heater type

Etched foil

Etched foil

Heater power

33W

33W

Warm up time

4 min approximately

4 min approximately

(ambient to 105°C)

(ambient to 105°C)

The heated lid will only come on
if the set temperature is above 35°C

Programming

Program naming using alpha keys

Yes

Yes

Program password protection

Yes

Yes

Number of programs (3 step)

80

80

Maximum number of steps per stage

35

35

Maximum number of stages per program

25

25

Minimum hold time

1 sec

1 sec

Maximum hold time

18 hrs

18 hrs

Incremental/decremental temperature

Yes

Yes

Incremental/decremental step hold time

Yes

Yes

Programmable ramp rate (heat or cool)

Variable in 0.1°C/sec steps

Variable in 0.1°C/sec steps

Programmable ramp rate resolution

0.1°C

0.1°C

19

Advertising