Techne TC-3000X User Manual

Page 39

Advertising
background image

Use the "Alpha/numeric" keys to change the
values of the 'fst' line to the first
temperatures and/or times you require.

Press "Enter" after each field you change.

You will exit 'Edit Mode' when you reach the
end of the line.

Use the "Down Arrow" key to move onto
the 'lst' line.

Press "Enter" to edit the step.

Use the "Alpha/numeric" keys to change the
values of the 'lst' line to the final
temperatures and/or times you require.

Press "Enter" for each field you change.

You will exit 'Edit Mode' when you reach the
end of the line.

37

Number cycles

30

Step 94.0° C 00m30

max

fst 65°C 00m30

max

[ edit temperature ]

Step 94.0° C 00m30

max

fst 65.0° C 00m30 max

lst 55.0° C 00m30

[ , to edit ]

Step 94.0° C 00m30

max

fst 65.0° C 00m30

max

lst 55.0° C 1m30

[ edit time

]

Number cycles

30

Step 94.0° C 00m30 max

fst 55.0° C 00m30

max

[ , to edit ]

The step will change to 'fst', meaning "first",
and the software will insert another line after
it entitled 'lst', meaning "last".

Press "Enter" twice to go into edit mode.

When you run the program it will automatically increment or decrement in equal steps from the first time
and/or temperature to the last over the number of cycles in the stage. For example: if the first
temperature is 65°C and the last temperature is 55°C, over 30 cycles the temperature will decrease
0.33°C each cycle. Similarly, in the example shown, the time would increase from 30 seconds to 1 minute
30 seconds in 2 second increments.

[ to save ]

[ to delete step

]

[ ‘.’ to insert step ]

[ .For hrs/mins ]

Advertising