Figure 2-10 menu – calibration (cal) – EOTech CNVD-T (SU-232/PAS) Clip-On Night Vision Device - Thermal User Manual

Page 31

Advertising
background image

2-15

Activating the CAL menu item initiates a calibration of the

CNVD-T. After prolonged use, or after transitions from one

temperature extreme to another, a degradation of the

thermal image may be noticed. When this occurs, the

CNVD-T should be calibrated as follows:

a. Install the objective lens cover over the objective

lens to block out all available light.

b. Activate the CAL menu item as shown below.
c. The message “CALIBRATING” will appear as the

unit performs the calibration. Once complete, the

message will disappear and the CNVD-T will revert

to normal viewing mode.

d. Remove the objective lens cover to resume normal

viewing.

Figure 2-10 Menu – Calibration (CAL)

The CNVD-T is equipped with an internal sensor that detects

significant changes in temperature. When this condition

occurs, the message “RECAL” will appear in the eyepiece

display, indicating that the operator should perform a

calibration of the CNVD-T as described above.