Roll Down Calibration for CycleOps PowerBeam Pro Trainer
The CycleOps PowerBeam Pro resistance unit is calibrated at the factory. This enables it to ac-
curately measure power using similar technology as a PowerTap. Rolling resistance is calibrated
into each individual resistance unit, but can vary depending on tire type, tire pressure, clamping
pressure and other conditions. To obtain the most accurate results, follow these guidelines:

Perform the Manual Calibration described below.

Inflate tires to the rated pressure.

Attach the bicycle according to the PowerBeam Pro trainer user guide

Clamp the resistance unit to the rear tire according to the PowerBeam Pro trainer user

guide

For mountain bicycles, use smooth tires. Knobby tires are noisy and contribute to inac

curacies.

1.Before entering the calibration routine, warm up the PowerBeam Pro and tires by riding the system for five
minutes.

1. Press and release [MODE] until Menu mode is displayed.

3. Press [JOYSTICK] in and release to select Sensors.

4. Press [JOYSTICK] up/down to highlight RU Sensor. Press [JOYSTICK] in and release to select RU Sensor.

5. Press [JOYSTICK] up/down to highlight the active RU Sensor. Note: the active RU Sensor is indicated by a
hash mark next to the name in the RU Sensor list. Press [JOYSTICK] in and release to select the active RU Sensor.

6. Press [JOYSTICK] up/down to highlight Calibration. Press [JOYSTICK] up/down to select Calibration.

7. Press [JOYSTICK up/down to highlight Roll Down calibration. Press [JOYSTICK] in and release to select Roll
Down calibration.